PROMAX 3D: ADJUSTMENTS & CALIBRATIONS

3.2 Checking the connection between PCs and X-ray unit

Switch on the ProMax X-ray unit, Romexis PC and the Reconstruction PC.
ProMax checks
Figure 78
The ProMax sends a Reply packet in case the Ethernet link is up and running.
Reconstruction PC checks
Connections
• When the Romexis program is ready for the exposure and the ProMax X-ray unit is in
Technical Manual
Check that the Ethernet LINK LED on the Ethernet PCB is on.
Check that you can ping the ProMax IP address from the Romexis PC to verify that the
TCP/IP connection works (executing the command "ping <ProMax IP address>", e.g.
ping 192.168.0.130, see below).
Check that you can ping the Reconstruction PC IP address from the Romexis PC to verify
that the TCP/IP connection works.
Ready state check that:
- The 3D sensor head indicator light is on.
- The Ethernet and Sensor ACT LEDs on the Ethernet PCB are blinking.
- The Sensor LINK LED on the Ethernet PCB is blinking.
Check that you can ping the ProMax 3D sensor IP address from the Romexis PC when the
3D sensor head indicator light is on.