Replacing Pcbs; C-Arm Pcb - Planmeca ProMax Technical Manual

Hide thumbs Also See for ProMax:
Table of Contents

Advertisement

Chapter E - C-ARM AND IMAGING ARM

8 REPLACING PCBs

8.1 C-arm PCB

a)
Remove the C-arm upper and inner covers, the cassette carriage/sensor head covers and
tube head inner cover as described in section 5 "REMOVING THE COVERS" on page E-77.
b)
Disconnect all the cables that are connected to the C-arm PCB.
c)
Unscrew the C-arm PCB attachment screws with 2.5mm Allen key and remove the PCB (Fig.
131, 1).
d)
Install the new C-arm PCB in reverse order. Connect the cables as shown in Fig. 131.
Figure 131
E-96 Planmeca ProMax
J3
J1
J2
J3
J4
J5
J7
121-10-05
J9
C-Arm
C-Arm PCB
Angle Sensor
J10
CAM
J1
121-10-08
J27
CH4 CH5 CH6
J21
J22
J23
J24
J26
J19
J5
DIMAX2
PAN
M4x6 DIN 7984
J1
J6
J8
J2
J11
CH7 CH8
Aux.
J18
J20
Aux.
J25
REPLACING PCBs