Calibration Errors (7Xx - Planmeca ProMax Technical Manual

Hide thumbs Also See for ProMax:
Table of Contents

Advertisement

Chapter C - HELP & ERROR MESSAGES

3.6 Calibration errors (7xx)

Error
Explanation
code
E701
Missing calibration
E702
E703
E704
E705
Preheat calibration
E706
E710
Primary collimator
E711
C-14 Planmeca ProMax
Tube head filament definition not done
Tube head kV-calibration not done
Panoramic DEC not calibrated
Cephalometric DEC not calibrated
Problem during the preheat calibration.
Problem with the preheat calibration.
Primary y-collimator opening is too big.
3D primary x-collimator opening is too
wide.
ERROR MESSAGES
Comments
Perform the preheat calibration
again.
Perform the preheat calibration
again.
Calibrate the DEC-PAN.
Calibrate the DEC-CEPH.
The current in the tube head is
not high enough. Perform the
preheat calibration again.
The current in the tube head is
not raising on the voltage area.
Perform the preheat calibration
again. Check the functionality of
the tube head.
Adjust the lower blade upwards
or the upper blade downwards.
Adjust the left blade to right or the
right blade to left.