ADJUSTMENTS AND CALIBRATIONS
Chapter H - CEPHALOSTAT
Select i47.2 for Cephalometric program from the appearing window.
Figure 66
The system suggests default kV and mA values. Touch the Ready field.
Figure 67
NOTE
In case the contrast on the exposure is not satisfactory, change the values by touching
the arrow fields and take a test exposure.
The following window appears on the computer screen.
Figure 68
Technical manual