ADJUSTMENTS AND CALIBRATIONS

1.11 Adjusting the cephalometric beam vertical position

Select kilovolt and milliampere values high enough to enable the radiation beam to be seen in
a darkened room. The actual values will depend on how dark the room is.
To enter the calibration mode first touch the i field on the Main display. Select Technical cali-
brations (i600) from the list on the display. The password to the Technical calibrations mode
is asked when the Technical calibrations mode is entered for the first time after switching the
unit on. The password is 1701.
Figure 43
From the list on the i600 display that appears select Primary collimator calibration (i610).
Figure 44
Technical manual
Chapter H - CEPHALOSTAT
Planmeca ProMax H-27