Planmeca ProMax Technical Manual page 141

Hide thumbs Also See for ProMax:
Table of Contents

Advertisement

PANORAMIC BEAM AND PATIENT POSITIONING
MECHANISM ADJUSTMENT
Select Technical calibrations (i600) from the list on the display. The password to the Techni-
cal calibrations mode is asked when the Technical calibrations mode is entered for the first
time after switching the unit on. The password is 1701.
Figure 59
From the list on the i600 display that appears select Layer laser calibration (i660).
Figure 60
Technical Manual
Chapter E - C-ARM AND IMAGING ARM
Planmeca ProMax E-43