Chapter A - GENERAL & TECHNICAL DATA
c)
NOTE
Values in this item are normalized to the CTDI
X- ray tube voltage: 70 kV
X- ray tube voltage: 50 kV
d)
Maximum deviation from the values given in items a), b) and c) is ±20%.
Dose profile
Graphical presentation of the dose profile along a line z perpendicular to the tomographic
plane measured in the centre location of the head-dosimetry phantom:
z-direction = Beam height at rotation centre
Sensitivity profile
Sensitivity profile N/A - the resolution is substantially equal in z-direction.
Weighted CTDI
CTDI
volume scanned for the selected CT conditions of operation.
CTDI
Volume CTDI
CTDI
Geometric efficiency in Z-direction
Geometric efficiency in z-direction is 100%.
A-14 Planmeca ProMax
Deviation from the default CT scanner
conditions of operation
0
0
100
(VolumeComputedTomographyDoseIndex) describes the average dose over the total
w
= 7.15 (5.72) mGy
W
w
= CTDI
= 7.15 (5.72) mGy
vol
W
PROMAX 3D: CTDI VALUES
in item a) 2)
100 (peripheral, max)
Relative CTDI
0.34 (0.34)
0.05 (0.05)
x-direction = Dose
100 (peripheral, max)
Technical Manual