EMC INFORMATION
portable and mobile RF communications equipment and Planmeca ProMax X-ray unit
Planmeca ProMax X-ray unit is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of Planmeca ProMax X-ray unit can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF com-
munications equipment (transmitters) and the Planmeca ProMax X-ray unit as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
Technical Manual
Recommended separation distances between
W
0.01
0.1
1
10
100
Chapter A - GENERAL & TECHNICAL DATA
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz 800 MHz to 2.5 GHz
d
=
1.2 P
0.2
0.4
1.2
4.0
12.0
m
d
=
1.2 P
d
0.2
0.4
1.2
4.0
12.0
Planmeca ProMax A-23
=
2.3 P
0.3
0.7
2.4
8.0
24.0