Chapter I - PLANMECA PROCEPH
1.2 Removing the necessary covers
Remove the tube head cover (refer to section 2 "REMOVING THE HOUSINGS AND COV-
ERS" on page I-18).
Remove the telescopic column rear cover plates, refer to section 2.2 "Removing the tele-
scopic column rear cover plates" on page G-11.
Remove the cephalostat housings and covers, refer to section 2 "REMOVING THE HOUS-
INGS AND COVERS" on page I-18.
1.3 Parameter values and beam check exposure
Set C-arm rotation to 0, Y-collimator bottom to 404 and X-collimator left to 3520 as described
below.
To enter the calibration mode first touch the i-field on the Main display. Select Technical cali-
brations (i600) from the list on the display. The password to the Technical calibrations mode
is asked when the Technical calibrations mode is entered for the first time after switching the
unit on. The password is 1701.
Figure 3
I-2 ProMax X-ray unit
ADJUSTMENTS AND CALIBRATIONS
Technical manual