SERVICE SETTINGS

5.6 Configuration selftest

From the list of Service settings select Configuration selftest (i560).
Touch the Reset system configuration field if you wish to bring back the default exposure
values for the quick buttons that are shown on the Select kV/mA display.
The Clear exposure history field clears the Error history (i420) and Exposure statistics
(i430) counters.
Figure 43
NOTE
It is important to perform the Reset system configuration function in case a device(s)
has/have been added or removed from the X-ray unit.

5.7 Select configuration

NOTE
This option will be shown only if a 3D license has been activated on display i310.
On the Service settings display, touch Select configuration (i570).
A display will appear where you can select the ProMax, collimator and sensor type.
Figure 44
Technical Manual
Chapter B - INFORMATION DISPLAYS
Planmeca ProMax B-29