Chapter
H

1 ADJUSTMENTS AND CALIBRATIONS

NOTE
During the adjustments write down the selected parameter values. After the cepha-
lostat adjustment switch the X-ray unit off and on and then check that the new parame-
ter values are stored into the memory.

1.1 Required tools

Figure 1
Figure 2
Technical manual
CEPHALOSTAT
Ceph head support alignment tool (part number 00653126).
Ceph head support alignment laser (part number 10004499). Used to check the position
of the cephalostat head support.
Planmeca ProMax H-1