Adjusting The Cephalometric Beam Top Limit - Planmeca ProMax Technical Manual

Hide thumbs Also See for ProMax:
Table of Contents

Advertisement

Chapter I - PLANMECA PROCEPH

1.8 Adjusting the cephalometric beam top limit

Take a beam check exposure. Refer to the Planmeca Device Tool manual, publication num-
ber 10031558. A fade-out area of 2-3mm must be seen in all edges of the X-ray beam. Adjust
the beam position as follows.
To enter the calibration mode first touch the i-field on the Main display. Select Technical cali-
brations (i600) from the list on the display. The password to the Technical calibrations mode
is asked when the Technical calibrations mode is entered for the first time after switching the
unit on. The password is 1701.
Figure 18
From the list on the i600 display that appears select Primary collimator calibration (i610).
Figure 19
I-14 ProMax X-ray unit
ADJUSTMENTS AND CALIBRATIONS
Technical manual

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Promax 3dPromax 3d s

Table of Contents