ERROR MESSAGES
Error
Explanation
code
E230
Overcurrent
detected in
E231
E232
E233
E234
E235
E236
E237
E238
E239
E240
E241
Technical Manual
0 - lift motor
1 - shoulder motor
2 - elbow motor
3 - temple rest motor
4 - Dimax3 CEPH scan motor (or CEPH
marking motor, depending on version)
5 - C-arm rotation motor
6 - receptor pan motor
7 - X-collimator motor
8 - Y-collimator motor
9 - receptor rotation motor
10 - receptor lift motor (reserved option)
11 - tube head lift motor (reserved option)
Chapter C - HELP & ERROR MESSAGES
Comments
Too high current detected in the
motor and the movement has
been stopped. Check the motor
cable. If necessary then replace
the PSU PCB.
Too high current detected in the
motor and the movement has
been stopped. Check the motor
cable.
Too high current detected in the
motor and the movement has
been stopped. Check the motor
cable.
Too high current detected in the
motor and the movement has
been stopped. Check the motor
cable.
Too high current detected in the
motor and the movement has
been stopped. Check the motor
cable.
Too high current detected in the
motor and the movement has
been stopped. Check the motor
cable.
Too high current detected in the
motor and the movement has
been stopped. Check the motor
cable.
Too high current detected in the
motor and the movement has
been stopped. Check the motor
cable.
Too high current detected in the
motor and the movement has
been stopped. Check the motor
cable.
Too high current detected in the
motor and the movement has
been stopped. Check the motor
cable.
Planmeca ProMax C-5
Need help?
Do you have a question about the ProMax and is the answer not in the manual?