3.2 X-ray generation related errors (3xx)

Error
Explanation
code
E301
Filament voltage
E302
E303
E304
E305
Technical Manual
0 - lift motor
1 - shoulder motor
2 - elbow motor
3 - temple rest motor
4 - Dimax3 CEPH scan motor (or CEPH
marking motor, depending on version)
5 - C-arm rotation motor
6 - receptor pan motor
7 - X-collimator motor
8 - Y-collimator motor
9 - receptor rotation motor
10 - receptor lift motor (reserved option)
11 - tube head lift motor (reserved option)
13 - beam limiting device
missing completely
too low during preheat
too high during preheat
dropped suddenly during exposure
cannot be enabled due to an internal
error
Chapter C - HELP & ERROR MESSAGES
Comments
Check the sensor, cable, motor
and the PSU PCB.
Check the sensors, cables and
the motor.
Check the sensors, cables and
the motor.
Check the sensors, cables and
the motor.
Check the sensors, cables and
the motor.
Check the sensors, cables and
the motor.
Check the sensors, cables and
the motor.
Check the sensors, cables and
the motor.
Check the sensors, cables and
the motor.
Check the sensors, cables and
the motor.
Check the adjustments and the
movement area.
Comments
Exposure not possible. Check the
cabling, interfaces and PSU-PCB
and tube head.
Exposure not possible. Check the
cabling, interfaces and PSU PCB
and tube head.
Exposure not possible. Check the
tube head and the PSU PCB.
Exposure interrupted. Check the
tube head and the PSU PCB.
Exposure interrupted. Check the
tube head and the PSU PCB.
Planmeca ProMax C-7