ERROR MESSAGES
Error
Explanation
code
E331
Arching
E332
E333
E334
E335
E336
E341
Monitoring errors
Technical Manual
across X-ray tube, minor
across X-ray tube, severe
at tube head anode end, minor
at tube head anode end, severe
at tube head cathode end, minor
at tube head cathode end, severe
kV-signal (exposure time) too short/long
Chapter C - HELP & ERROR MESSAGES
Comments
Code not displayed.
Exposure interrupted. Check the
functionality of the tube head.
If the X-ray unit has not been
used for a week or more, enter
display i460 and take four test
exposures with a one-minute
interval between exposures. Use
the lowest possible kV and mA
values and a long exposure time.
Code not displayed.
Exposure interrupted.
Code not displayed.
Exposure interrupted.