SERVICE SETTINGS
5.4 Exposure program tuning
From the list of Service settings select Exposure program tuning (i540). In this mode the
maximum and minimum values of the kV and the minimum mA range can be set for pano-
ramic, cephalometric and 3D exposures.
Figure 39
To set the panoramic exposure values select Set pan parameters (i54.1) limits. The display
shown below appears. Select the maximum and minimum kilovolt, and minimum milliam-
pere values with the arrow fields and touch the Done field.
Figure 40
Technical Manual
Chapter B - INFORMATION DISPLAYS
Planmeca ProMax B-27