Chapter H - CEPHALOSTAT
To calibrate the Cephalometric DEC select Cephalometric DEC (i632).
Figure 92
The Cephalometric DEC display appears. The exposure values suggested on this display are
66 kV value and 5 mA. Use these values when taking an OD exposure. Touch the OD
exposure field.
Figure 93
H-52 Planmeca ProMax
ADJUSTMENTS AND CALIBRATIONS
Technical manual