1.8 Cephalostat head support position adjustment

Adjusting the ear posts in horizontal direction
Switch the unit off. Remove the cephalostat housings according to the instructions given in
section 4 "REMOVING THE HOUSINGS AND COVERS" on page H-64.
Loosen the locking screw of the adjustment flange. Rotate the head support around its verti-
cal axis to adjust the ear posts in horizontal direction as follows.
Loosen the head support attachment screws (Fig. 27, 1) and the alignment bar screw (Fig.
27, 2). Rotate the alignment bar until the head support is in correct position (Fig. 27, 3 and 4).
4
Figure 27
H-16 Planmeca ProMax
Locking screw of the adjustment flange
Adjusting the ear posts in horizontal direction