ADJUSTING AND CALIBRATING X-RAY
BEAM
2.4

Step 1: Adjusting the X-ray beam position

To be able to check the beam position without the C-arm moving enter the collimator
calibration mode as follows:
First touch the i-field on the main display. The main display is the display that is shown
when the X-ray unit is switched on. The Configuration display will appear. On the
Configuration display, select the option Collimator calibration. The Collimator calibration
display will appear.
NOTE You will need a password to gain access to the information displays. The password
needs to be entered each time the X-ray unit is switched on. The password is 1701.
Figure 13
Technical Manual
CHAPTER C
C-ARM AND UPPER ARM
Planmeca ProOne C-9