Performance Tests

11. Receiver Input Impedance

11. Receiver Input Impedance
Perform this test to verify the return loss of each of the analyzer receiver inputs.
Analyzer warmup time: 30 minutes
Specifications
Frequency Range
300 kHz–2 MHz
2 MHz–1.3 GHz
1.3 GHz–3 GHz
a
3 GHz–6 GHz
a. For analyzers with Option 006.
Required Equipment
Description
Test set, 50 Ω
2-way power splitter, 50 Ω
Attenuator, 10 dB
Termination, 50 Ω
Calibration kit, 50 Ω, type-N
Adapter, APC-7 to Type-N (m)
RF cable set, 50 Ω, Type-N
RF cable, 50 Ω, 7mm
Procedure
1. Connect the equipment as shown in
• If you are using the HP/Agilent 85044A, set the attenuation to 0 dB.
• If you are using the HP/Agilent 85046A or 85047A S-Parameter test set, connect the
test set interface cable to the analyzer and use test set port 1.
2-42
Receiver Input
Impedance
≥ 20 dB return loss
≥ 24 dB return loss
≥ 19 dB return loss
≥ 15 dB return loss
HP/Agilent Part or Model Number
85044A/85046A/85047A
11667A Option 001
8491A Option 010
part of cal kit 85032B
85032B
11525A
11851B
part of 11857D
Figure
2-18, with the following additions: