Performance Tests
8. Receiver Phase Frequency Response (Ratio)
Figure 2-13
Phase Frequency Response Test Setup
From 300 kHz to 3 GHz
22.Press
CONTINUOUS
23.Press
IF BW
Avg
24.Press
Meas
A/B
25.Only For Analyzers with Option 006: Press
26.Press
Scale Ref
knob to vary the electrical delay until the trace is in the most linear horizontal position.
27.Press
Sweep Setup
SEARCH: MAX
Marker Search
. Record the marker 1 or marker 2 value—whichever has the greatest
SEARCH: MIN
absolute value—in the "Measured Value" column of the performance test record.
From 3 GHz to 6 GHz (Option 006)
28.Press
CONTINUOUS
29.Press
Scale Ref
to vary the electrical delay until the trace is in the most linear horizontal position.
30.Press
Sweep Setup
SEARCH: MAX
Marker Search
. Record the marker 1 or marker 2 value—whichever has the greatest
SEARCH: MIN
absolute value—in the "Measured Value" column of the performance test record.
2-30
Start
300
k/m
.
300
x1
.
Format
PHASE
ELECTRICAL DELAY
0.6
x1
TRIGGER MENU
. Press
Start
3
G/n
5
x1
ELECTRICAL DELAY
TRIGGER MENU
. Press
.
Stop
3
G/n
. Turn the analyzer front panel
. Press
SINGLE
Marker
MARKER 2
Marker
.
Stop
6
G/n
. Turn the analyzer front panel knob
. Press
SINGLE
Marker
MARKER 2
Marker
.
MARKER 1
Marker Search
MARKER 1
Marker Search