Agilent Technologies 8753ES Option 011 Service Manual page 123

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Figure 3-2
Setup A for RF Output Correction Constants
10.Press
Sweep Setup
11.Record the power meter reading in the first column of
Table 3-2
Power Meter Readings
Setup A Reading
(First Reading)
300 kHz: _______dB
50 MHz: _______dB
1.5 GHz: _______dB
3 GHz:_______ dB
a
6 GHz
: _______dB
a. For Option 006 instruments only.
12.Repeat the previous two steps at 50 MHz, 1.5 GHz, and 3 GHz.
13.For Option 006 Instruments Only: Make a measurement at 6 GHz by disconnecting
the 8482A (sensor A) from the power splitter, and replacing it with the 8481A (sensor
B).
• If you are using the 438A power meter, the 8481A should be connected to the meter's
channel B input.
• If you are using the 437B power meter, zero and calibrate the 8481A sensor.
14.Reconfigure the equipment as shown in
Chapter 3
CW FREQ
300
k/m
Setup B Reading
(Second Reading)
minus _______dB
minus _______dB
minus _______dB
minus _______dB
minus _______dB
Figure
Adjustments and Correction Constants
RF Output Power Correction Constants (Test 47)
.
Table
3-2.
Power Loss of #2
(Enter in Analyzer)
equals _______dB
equals _______dB
equals _______dB
equals _______dB
equals _______dB
3-3.
3-13