CURRENT PARAMETER NOT IN CAL SET
Error Number 64
Correction is not valid for your selected measurement parameter. Either change the
measurement parameters or perform a new calibration.
DEADLOCK
Error Number 111
A fatal firmware error occurred before instrument preset completed.
DEVICE: not on, not connect, wrong addrs
Error Number 119
The device at the selected address cannot be accessed by the analyzer. Verify that the
device is switched on, and check the GPIB connection between the analyzer and the
device. Ensure that the device address recognized by the analyzer matches the GPIB
address set on the device itself.
DISK HARDWARE PROBLEM
Error Number 39
The disk drive is not responding correctly. Refer to the disk drive operating manual.
DISK MESSAGE LENGTH ERROR
Error Number 190
The analyzer and the external disk drive aren't communicating properly. Check the
GPIB connection and then try substituting another disk drive to isolate the problem
instrument.
DISK: not on, not connected, wrong addrs
Error Number 38
The disk cannot be accessed by the analyzer. Verify power to the disk drive, and check
the GPIB connection between the analyzer and the disk drive. Ensure that the disk
drive address recognized by the analyzer matches the GPIB address set on the disk
drive itself.
DISK READ/WRITE ERROR
Error Number 189
There may be a problem with your disk. Try a new floppy disk. If a new floppy disk
does not eliminate the error, suspect hardware problems.
INITIALIZATION FAILED
Error Number 47
The disk initialization failed, probably because the disk is damaged.
INSUFFICIENT MEMORY, PWR MTR CAL OFF
Error Number 154
There is not enough memory space for the power meter calibration array. Increase the
available memory by clearing one or more save/recall registers, or by reducing the
number of points.
NO CALIBRATION CURRENTLY IN PROGRESS
Error Number 69
The
already in progress. Start a new calibration.
Chapter 10
RESUME CAL SEQUENCE
Service Key Menus and Error Messages
softkey is not valid unless a calibration is
Error Messages
10-45