Agilent Technologies 8753ES Option 011 Service Manual page 348

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Error Terms
Full Two-Port Error-Correction Procedure
Error Term Inspection
If the correction is not active, press
NOTE
1. Press
SERVICE MENU
System
The analyzer copies the first calibration measurement trace for the selected error term
into memory and then displays it.
2. Press
Scale Ref
3. Press
Marker Fctn
magnitude.
4. Compare the displayed measurement trace to the traces shown in
Descriptions" on page
5. Make a hardcopy of the measurement results:
a. Connect a printing or plotting peripheral to the analyzer.
b. Press
SYSTEM CONTROLLER
Local
appropriate peripheral to verify that the GPIB address is set correctly on the
analyzer.
c. Press
Save/Recall
d. Press
Display
later.
For detailed information on creating hardcopies, refer to your analyzer's user's guide.
11-8
TESTS
Table 11-1
and adjust the scale and reference to study the error term trace.
and use the marker functions to determine the error term
11-9, and to previously measured data.
and then choose either
and title each data trace so that you can identify it
MORE
TITLE
Cal
CORRECTION ON
EXECUTE TEST
32
x1
lists the test numbers.
and select the
SET ADDRESSES
or
PRINT
PLOT
.
.
"Error Term
.
Chapter 11