B/R Phase Frequency Response
12.Connect the second type-N cable from the RF cable set to port B.
From 300 kHz to 3 GHz
13.Press
CONTINUOUS
14.Press
B/R
Meas
15.Only For Analyzers with Option 006: Press
16.Press
Scale Ref
knob to vary the electrical delay until the trace is in the most linear horizontal position.
17.Press
Sweep Setup
SEARCH: MAX
Marker Search
. Record the marker 1 or marker 2 value—whichever has the greatest
SEARCH: MIN
absolute value—in the "Measured Value" column of the performance test record.
From 3 GHz to 6 GHz (Option 006)
18.Press
CONTINUOUS
19.Press
Scale Ref
to vary the electrical delay until the trace is in the most linear horizontal position.
20.Press
Sweep Setup
MARKER 1 Marker Search
MARKER 2 Marker Search
value—whichever has the greatest absolute value—in the "Measured Value" column of
the performance test record.
A/B Phase Frequency Response
21.Connect the equipment as shown in
Chapter 2
Start
300
k/m
.
PHASE
Format
ELECTRICAL DELAY
0.6
x1
TRIGGER MENU
. Press
Start
3
G/n
ELECTRICAL DELAY
5
x1
TRIGGER MENU
SEARCH: MAX
SEARCH: MIN
Figure
8. Receiver Phase Frequency Response (Ratio)
.
Stop
3
G/n
. Turn the analyzer front panel
. Press
SINGLE
Marker
MARKER 2
Marker
.
Stop
6
G/n
. Turn the analyzer front panel knob
. Press
SINGLE
Marker
. Press
Marker
. Record the marker 1 or marker 2
2-13.
Performance Tests
.
MARKER 1
Marker Search
2-29

Advertisement