Digital Control - Agilent Technologies 8753ES Option 011 Service Manual

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Start Troubleshooting Here

Digital Control

Digital Control
Observe the Power Up Sequence
Switch the analyzer power off, then on. The following should take place within a few
seconds:
• On the front panel, observe the following:
1. All six amber LEDs illuminate.
2. The amber LEDs go off after a few seconds, except the CH1 LED. See
• The display should come up bright with no irregularity in colors.
• After an initial pattern, five red LEDs on the A9 CPU board should remain off. They can
be observed through a small opening in the rear panel.
If the power up sequence does not occur as described, or if there are problems using the
front panel keyboard, refer to
Figure 4-5
Front Panel Power Up Sequence
4-12
Chapter 6 , "Digital Control Troubleshooting."
Figure
4-5.
Chapter 4

Advertisement