Agilent Technologies 8753ES Option 011 Service Manual page 131

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Figure 3-6
Input R Sampler Correction Setup
12.Press
CONTINUE
The analyzer starts the first part of the automatic adjustment. This part will take about
seven minutes.
13.For Option 006 Instruments Only: After the analyzer has finished the first part of
the adjustment, disconnect the 8482A (sensor A) from the power splitter, and replace it
with the 8481A (sensor B) for the 6 GHz measurement.
• If you are using the 438A power meter, the 8481A should be connected to the meter's
channel B input.
• If you are using the 437B power meter, zero and calibrate the 8481A sensor.
14.Press
CONTINUE
15.Connect the equipment as shown in
at the power splitter outputs.
Chapter 3
.
.
Figure
Adjustments and Correction Constants
Sampler Magnitude and Phase Correction Constants (Test 53)
3-7. Use two cables of equal electrical length
3-21