In Case of Difficulty
1. Ensure that the power meter and power sensor are operating to specification.
2. Inspect the power splitter connectors. Poor match at these connections can generate
power reflections that can cause the analyzer to appear to be out of specification.
3. Inspect the analyzer RF OUT connector for damage.
4. If any test fails, perform
page
3-11.
5. For a source failure, refer to
information.
Chapter 2
"RF Output Power Correction Constants (Test 47)" on
Chapter 7 , "Source Troubleshooting,"
2. Source Power Range, Linearity, and Accuracy
for troubleshooting
Performance Tests
2-11