Performance Tests

13. Receiver Magnitude Compression

13. Receiver Magnitude Compression
Perform this test to verify the compression/expansion magnitude levels of the analyzer's
receiver samplers. Power sweeps from low to high power are made at designated CW
frequencies. A reference measurement is made while the signal to the receiver is
attenuated to avoid compression. The attenuation is removed and compression is observed
and measured.
Analyzer warmup time: 30 minutes
Specifications
Frequency Range
300 kHz–3 GHz
b
3 GHz–6 GHz
300 kHz–3 GHz
b
3 GHz–6 GHz
a. With 10 Hz IF bandwidth.
b. Only for analyzers with Option 006.
Required Equipment
Description
Power meter
Power sensor
Attenuator, 3 dB
Attenuator, 10 dB
Attenuator, 20 dB
Cable (2), 50Ω type-N, 24-inch
2-way power splitter (2), 50Ω
Additional Equipment for Option 006 Analyzers
Power sensor
Channel A Magnitude Compression
1. Zero and calibrate the power meter.
2. Connect the equipment as shown in
2-58
Receiver Channel
Magnitude
A
A
B
B
HP/Agilent Part or Model Number
437B/438A or E4418B/E4419B
8482A
8491A Option 003
8491A Option 010
8491A Option 020
8120-4781
11667A Option 001
8481A
Figure
a
≤ 0.55 dB
≤ 0.87 dB
≤ 0.55 dB
≤ 0.87 dB
2-27.
Chapter 2