Agilent Technologies 8753ES Option 011 Service Manual page 87

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Procedure
A Channel Harmonic Measurement Accuracy
1. Connect the equipment as shown in
Figure 2-33
Harmonic Measurement Accuracy Test Setup
2. Zero and calibrate the power meter. Set the power meter to measure dBm. Refer to the
power meter operating manual for more information on how to perform these tasks.
3. Preset the external source, and set the power level to −3.5 dBm and the CW frequency
to 32 MHz. Set the power meter for the appropriate calibration factor.
4. Write the power meter reading on the performance test record. It should read
approximately −10 dBm.
5. On the network analyzer, press
the analyzer internal source to 16 MHz.
6. Press
System
HARMONIC MEAS
7. Press
Marker
Meas
display, in the "Input A Value" column on the performance test record.
8. On the external source, set the CW frequency to 48 MHz. Reset the power meter for the
appropriate calibration factor.
9. Record the power meter reading on the performance test record.
Chapter 2
Figure
2-33.
Preset
Sweep Setup
SECOND
. Write the marker 1 reading, which appears on the analyzer
A
16. Harmonic Measurement Accuracy (Option 002 Only)
CW FREQ
.
Performance Tests
to set
16
M/µ
2-71