Agilent Technologies 8753ES Option 011 Service Manual page 330

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Service Key Menus and Error Messages
Service Key Menus
PEEK/POKE Menu
To access this menu, press
Table 10-13
PEEK/POKE Menu Keys
Key
PEEK/POKE
PEEK/POKE ADDRESS
PEEK
POKE
RESET MEMORY
10-40
SERVICE MENU
System
GPIB
Mnemonic
N/A
Allows you to edit the content of one or more memory
addresses. The keys are described below.
CAUTION
Accesses any memory address and shows it in the active
PEEL[D]
entry area of the display. Use the front panel knob, entry
keys, or step keys to enter the memory address of interest.
Displays the data at the accessed memory address.
PEEK
allows you to change the data at the memory address
POKE[D]
accessed by the
the front panel knob, entry keys, or step keys to change the
data. The A9CC switch must be in the "ALTER" position in
order to poke.
N/A
Resets or clears the memory where instrument states are
stored. To do this, press
.
PEEK/POKE
Description
The PEEK/POKE capability is intended for
service use only.
PEEK/POKE ADDRESS
RESET MEMORY
softkey. Use
Preset
.
Chapter 10

Advertisement

Table of Contents
loading