Error Terms
Full Two-Port Error-Correction Procedure
c. Press
Cal
REV ISOL'N ISOL'N STD
d. Return the averaging to the original state of the measurement, and press
RESUME CAL SEQUENCE
17.To compute the error coefficients, press
The analyzer displays the corrected measurement trace. The analyzer also shows the
notation Cor at the left of the screen, indicating that error-correction is on.
You can save or store the measurement correction to use for later
NOTE
measurements. Use the menus under
user's guide for procedures.
18.This completes the full two-port correction procedure. You can connect and measure
your device under test.
11-6
RESUME CAL SEQUENCE
ISOLATION DONE
.
FWD ISOL'N ISOL'N STD
.
DONE 2-PORT CAL
, or refer to your analyzer's
Save/Recall
Cal
.
Chapter 11