Agilent Technologies 8753ES Option 011 Service Manual page 69

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

11.Take a power meter calibration sweep.
Cal
PWRMTR CAL
ONE SWEEP
12.Verify that the power meter reads approximately −30 dBm.
Full 2-Port Calibration
13.Connect the equipment as shown in the
the B channel connectors.
Figure 2-23
Full 2-Port Calibration
14.Perform a full 2-port error correction with isolation.
15.Save the results of the new cal set.
Measure Channel B Magnitude Dynamic Accuracy
16.Connect the equipment as shown in
to 20 dB.
Chapter 2
−30
x1
TAKE CAL SWEEP
Figure
12. Receiver Magnitude Dynamic Accuracy
Figure
2-23. Be sure to reconnect the cable to
2-24. Confirm that the step attenuator is set
Performance Tests