Agilent Technologies 8753ES Option 011 Service Manual page 55

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Figure 2-17
Receiver Trace Noise Test Setup
2. Press
Preset
Power
3. Press
CW FREQ
4. Press
Marker Fctn
marker statistics feature.
A/R Trace Noise Magnitude IF BW = 10 Hz
5. Press
Avg
IF BW
6. Press
A/R
Meas
7. Press
Sweep Setup
indicated by the Hld notation on the left side of the display.
8. Press
Scale Ref
9. Write the s. dev (standard deviation) value shown, which appears on the analyzer
display, on the performance test record.
A/R Trace Noise Phase IF BW = 10 Hz
10.Press
PHASE
Format
11.Press
Sweep Setup
indicated by the Hld notation on the left side of the display.
12.Press
Scale Ref
13.Write the s. dev (standard deviation) value shown, which appears on the analyzer
display, on the performance test record.
Chapter 2
10
x1
Sweep Setup
.
3
G/n
MARKER MODE MENU
.
10
x1
.
TRIGGER MENU
.
AUTO SCALE
.
TRIGGER MENU
.
AUTO SCALE
NUMBER of POINTS
MKR STATS ON
. Wait for the sweep to finish as
SINGLE
. Wait for the sweep to finish as
SINGLE
Performance Tests
10. Receiver Trace Noise
.
51
x1
to enable the analyzer
2-39