Performance Tests
15. Source and Receiver Harmonics (Option 002 Only)
If you have Option 006, press
12.Press
HARMONIC MEAS
System
13.After one sweep, press
trace.
14.Press
Scale Ref
15.Press
HARMONIC MEAS
System
16.After one sweep, press
17.Press
Marker Search
18.Write the marker 1 value on the performance test record.
Receiver Harmonics
19.Connect the equipment as shown in
Figure 2-32
Receiver Harmonics Test Setup
20.Press
Power
6
frequency range:
If you do not have Option 006, press
If you have Option 006, press
21.Repeat steps 5 through 10 to measure the receiver worst case 2nd harmonic for the
receiver input B. Press
2-68
Stop
2
HARMONIC OFF
DATA → MEMORY
Display
to get a better viewing of the trace.
AUTO SCALE
HARMONIC THIRD
Scale Ref
AUTO SCALE
.
SEARCH:MAX
Figure
x1
System
HARMONIC MEAS
Stop
Stop
3
in step 5.
Meas
B
.
G/n
.
DATA/MEM
.
.
2-32.
HARMONIC OFF
.
1.5
G/n
.
G/n
to normalize the
. To set the
Chapter 2