Performance Tests
7. Receiver Magnitude Frequency Response
Figure 2-9
Magnitude Frequency Response Test Setup (Receiver Input R)
2. Zero and calibrate the power meter. Set it to measure dBm.
3. Press
Preset
Marker
4. Press
Sweep Setup
meter for this CW frequency.
5. Write the power meter reading on the performance test record.
6. Write the marker reading, which appears on the analyzer display, in the "R Input
Power" column of the performance test record.
7. Repeat steps 4, 5 and 6 for the other CW frequencies listed on the performance test
record.
Input A Magnitude Frequency Response
8. Connect the equipment as shown in
2-24
R
Meas
Power
CW FREQ
300
k/m
Figure
.
6
x1
. Set the calibration factor on the power
2-10.
Chapter 2