Agilent Technologies 8753ES Option 011 Service Manual page 23

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Procedure
Path Loss Calibration
1. Connect the equipment as shown in
Figure 2-2
Path Loss Calibration Test Setup (#1)
2. Zero and calibrate the power meter. Also set the power meter for dBm. For more
information on how to perform this task, refer to the power meter documentation.
3. Press
Preset
Power
4. Press
Sweep Setup
this CW frequency (except if you are using an HP/Agilent 436A).
5. Write the power meter reading in the "First Value" column on the performance test
record.
6. Repeat steps 4 and 5 for the other CW frequencies listed on the performance test record.
7. Connect the equipment as shown in
Chapter 2
Figure
.
10
x1
CW FREQ
300
k/m
Figure
2. Source Power Range, Linearity, and Accuracy
2-2.
. Set the power meter calibration factor for
2-3.
Performance Tests
2-7

Advertisement

Table of Contents
loading

Table of Contents