4. Press
CONTINUE
• If DONE is displayed, you have completed this procedure.
• If FAIL is displayed, check that the RF cable is connected between Port 1 and Port 2.
Then, repeat this adjustment routine.
• If the analyzer continues to fail the adjustment routine, refer to
Control Troubleshooting."
Chapter 3
and observe the analyzer for the results of the adjustment routine:
Adjustments and Correction Constants
IF Amplifier Correction Constants (Test 51)
Chapter 6 , "Digital