Error Messages - Agilent Technologies 8753ES Option 011 Service Manual

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Service Key Menus and Error Messages

Error Messages

Error Messages
This section contains an alphabetical list of the error messages that pertain to servicing
the analyzer. The information in the list includes explanations of the displayed messages
and suggestion to help solve the problem.
NOTE
The error messages that pertain to measurement applications are included in
the your analyzer's reference guide.
BATTERY FAILED. STATE MEMORY CLEARED
Error Number 183
The battery protection of the non-volatile SRAM memory has failed. The SRAM
memory has been cleared. Refer to
Post-Repair Procedures,"
titled "Preset State and Memory Allocation" in the analyzer's reference guide for more
information about the SRAM memory.
BATTERY LOW! STORE SAVE REGS TO DISK
Error Number 184
The battery protection of the non-volatile SRAM memory is in danger of failing. If this
occurs, all of the instrument state registers stored in SRAM memory will be lost. Save
these states to a disk and refer to
Post-Repair Procedures,"
titled "Preset State and Memory Allocation" in the analyzer's reference guide for more
information about the SRAM memory.
CALIBRATION ABORTED
Error Number 74
You have changed the active channel during a calibration so the calibration in progress
was terminated. Make sure the appropriate channel is active and restart the
calibration.
CALIBRATION REQUIRED
Error Number 63
A calibration set could not be found that matched the current stimulus state or
measurement parameter. You will have to perform a new calibration.
CORRECTION CONSTANTS NOT STORED
Error Number 3
A store operation to the EEPROM was not successful. You must change the position of
the switch on the A9 CPU assembly. Refer to
CORRECTION TURNED OFF
Error Number 66
Critical parameters in your current instrument state do not match the parameters for
the calibration set, therefore correction has been turned off. The critical instrument
state parameters are sweep type, start frequency, frequency span, and number of
points.
10-44
Chapter 14 , "Assembly Replacement and
for battery replacement instructions. See the chapter
Chapter 14 , "Assembly Replacement and
for battery replacement instructions. See the chapter
"A9 Switch Positions" on page 3-6
.
Chapter 10