Performance Tests
16. Harmonic Measurement Accuracy (Option 002 Only)

16. Harmonic Measurement Accuracy

(Option 002 Only)

This test verifies the network analyzer's accuracy when operating in the harmonic
measurement mode (Option 002). The analyzer's reading is compared to that of the power
meter. The allowable difference is shown in the "Specifications" table.
Analyzer warmup time: 30 minutes
Specifications
Frequency Range
16 MHz–3 GHz
b
3 GHz– 6 GHz
a. At 25 °C ± 5 °C, A, B, R, −10 dBm input.
b. Only for analyzers with Options 002 and 006.
Required Equipment
Description
External source
Power meter
Power sensor
2-way power splitter, 50Ω
Attenuator, 10 dB
RF cable set, 50Ω type-N
Cable, BNC 24-inch
Adapter, 3.5-mm (f) to type-N (f)
Additional Equipment for Option 006 Analyzers
Power sensor
2-70
Harmonic Measurement
a
Accuracy
± 1.5 dB
± 3 dB
HP/Agilent Part or Model Number
83620A or 83712A
436A/437B/438A or E4418B/E4419B
8482A
11667A Option 001
8491A Option 010
11851B
8120-1840
1250-1745
8481A
Chapter 2