Loading Firmware

The following procedures will load firmware for new or existing CPU boards in your
network analyzer.
Analyzer warmup time: None required.
Required Equipment and Tools
Firmware disk for the 8753ES
Loading Firmware into an Existing CPU
Use this procedure for upgrading firmware in an operational instrument whose CPU board
has not been changed.
Loading firmware will clear all internal memory. Perform steps 1 through 5,
CAUTION
next, to save any instrument states that are stored in internal memory to a
floppy disk. If saving states is not necessary, proceed to step 6.
1. Press
Save/Recall
2. Select an instrument state and press
3. Press
SELECT DISK
4. If the instrument state file was not saved to disk with the same name that it had while
in internal memory, you may wish to rename the file:
Press
FILE UTILITIES
5. Repeat steps 1 through 4 for each instrument state that you wish to save.
6. Turn off the network analyzer.
7. Insert the firmware disk into the instrument's disk drive.
8. Turn the instrument on. The firmware will be loaded automatically during power-on.
The front panel LEDs should step through a sequence as firmware is loaded. The
display will be blank during this time.
At the end of a successful loading, the LEDs for Channel 1 and Testport 1 will remain
on and the display will turn on indicating the version of firmware that was loaded.
Chapter 3
SELECT DISK
INTERNAL MEMORY
RECALL STATE
INTERNAL DISK
RENAME FILE
Adjustments and Correction Constants
RETURN
.
RETURN
SAVE STATE
, enter the desired name, and press
Loading Firmware
.
.
.
DONE
3-35