Adjustments and Correction Constants

Unprotected Hardware Option Numbers Correction Constants

Unprotected Hardware Option Numbers Correction
Constants
Analyzer warmup time: None.
This procedure stores the instrument's unprotected option(s) information in A9 CPU
assembly EEPROMs.
1. Make sure the A9 switch is in the Alter position.
2. Record the installed options that are printed on the rear panel of the analyzer.
3. Press
SERVICE MENU
System
4. Refer to
Table 3-3
address for the specific installed hardware option that needs to be enabled or disabled.
Follow the address entry by
• Pressing
POKE
• Pressing
POKE
5. Repeat steps 3 and 4 for all of the unprotected options that you want to enable.
6. After you have entered all of the instrument's hardware options, press the following
keys:
SERVICE MENU
System
7. View the analyzer display for the listed options.
8. When you have entered all of the hardware options, return the A9 switch to the Normal
position.
9. Perform the
"EEPROM Backup Disk Procedure" on page
In Case of Difficulty
If any of the installed options are missing from the list, return to step 2 and re-enter the
missing option(s).
3-52
PEEK/POKE
for the address of each unprotected hardware option. Enter the
−1
POKE
−1
after an entry enables the option.
x1
after an entry disables the option.
0
x1
Table 3-3
Hardware
Options
1D5
011
FIRMWARE REVISION
PEEK/POKE ADDRESS
.
x1
PEEK/POKE Addresses
PEEK/POKE
Address
1619001529
1619001532
3-33.
.
Chapter 3