Agilent Technologies 8753ES Option 011 Service Manual page 297

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Edit List Menu
To access this menu, press
LOSS/SENSR LISTS
CAL FACTOR SENSOR B
Table 10-4
Edit List Menu Keys
Key
SEGMENT
EDIT
DELETE
ADD
CLEAR LIST
DONE
Self Diagnose Softkey
You can access the self diagnosis function by pressing
. This function examines, in order, the pass/fail status of all internal
SELF DIAGNOSE
tests and displays NO FAILURE FOUND if no tests have failed.
If a failure is detected, the routine displays the assembly or assemblies most probably
faulty and assigns a failure probability factor to each assembly.
Chapter 10
SERVICE MENU
System
, and then press one of the following:
or
POWER LOSS
GPIB
Mnemonic
N/A
Selects a segment (frequency point) to be edited, deleted
from, or added to the current data table. Works with the
entry controls.
Allows modification of frequency, cal factor and loss values
SEDI[D]
previously entered in the current data table.
Deletes frequency, cal factor and loss values previously
SDEL
entered in the current data table.
Adds new frequency, cal factor and loss values to the current
SADD
data table up to a maximum of 12 segments (frequency
points, PTS).
CLEL
Deletes the entire current data table (or list) when
pressed. Press
Selects whether or not to use the power loss table for a power
EDITDONE
meter calibration.
Service Key Menus and Error Messages
TEST OPTIONS
CAL FACTOR SENSOR A
.
Description
NO
to avoid deletion.
SERVICE MENU
System
Service Key Menus
or
YES
is
10-7