Check 11, Dc-Dc Converter Open Output Voltages; Dc-Dc Converter Output Connector; Dc-Dc Converter Open Output Voltages - Agilent Technologies 4349B Service Manual

4-channel high resistance meter
Hide thumbs Also See for 4349B:
Table of Contents

Advertisement

Check 11, DC-DC Converter Open Output Voltages

1. Turn o the 4349B.
2. Disconnect the DC-DC Converter Output Cable shown in Figure 4-12 from connector \J15"
on the A1 Main Board Assembly.
3. Turn the 4349B on.
4. Using a multimeter, check if the DC-DC Converter open output voltages on the output
connector pins shown in Figure 4-12 are in the ranges shown in Table 4-2.
5. If passed, replace the A1 Main Board Assembly, otherwise replace the DC-DC Converter
Assembly.
4-16
T roubleshooting
Figure 4-12. DC-DC Converter Output Connector
Table 4-2. DC-DC Converter Open Output Voltages
Pin (+) Pin ( ) Voltage Range
1
4
5
6
2
5.1
0.1 V
6
3
12
0.6 V
6
3
12
0.6 V

Advertisement

Table of Contents
loading

Table of Contents