Figure 9-3
Switch Positions on the A9 CPU
2. Press
Preset
System
.
x1
3. "Poke" the address for the appropriate test set
• 85047A: Press
• 85046A/B: Press
4. Measure the DC voltage at pin 14 (see
interconnect connector.
Figure 9-4
Test Set Interconnect Connector Pins
Chapter 9
SERVICE MENU
POKE
5
x1
Preset
POKE
1
x1
Preset
Figure
PEEK/POKE ADDRESS
.
.
9-4) of the analyzer rear panel test set
Accessories Troubleshooting
Test Set Troubleshooting
1619001527