Analog Bus Correction Constants (Test 46) - Agilent Technologies 8753ES Option 011 Service Manual

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Analog Bus Correction Constants (Test 46)

Analyzer warmup time: 30 minutes.
This procedure calibrates the analog bus by using three reference voltages (ground, +0.37
and +2.5 volts), then stores the calibration data as correction constants in EEPROMs.
1. Press
Preset
System
.
YES
2. Observe the analyzer for the results of the adjustment routine:
• If the analyzer displays ABUS Cor DONE, you have completed this procedure.
• If the analyzer displays ABUS Cor FAIL, refer to
Troubleshooting."
Chapter 3
SERVICE MENU
TESTS
Adjustments and Correction Constants
Analog Bus Correction Constants (Test 46)
46
x1
EXECUTE TEST
Chapter 6 , "Digital Control
3-9