Agilent Technologies 8753ES Option 011 Service Manual page 326

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Service Key Menus and Error Messages
Service Key Menus
Node 23
VCO Tune (A12 VCO tuning voltage)
Perform Step A12 and then press
Start
11
M/µ
Stop
21
M/µ
Meas
.
ANALOG IN
23
x1
Marker
Scale Ref
AUTO SCALE
The trace should show a voltage step as shown in
Figure
10-15. At normal operation, the
left half trace should be 0 ±1000 mV and the right half trace should be 100 to 200 mV
higher (that is, one to two divisions). If the trace does not appear as shown in
Figure
10-15,
refer to
"High/Low Band Transition Adjustment" on page
3-45.
Figure 10-15
Analog Bus Node 23
10-36