Source Frequency Range And Accuracy - Agilent Technologies 8753ES Option 011 Service Manual

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Performance Tests

1. Source Frequency Range and Accuracy

1. Source Frequency Range and Accuracy
Perform this test to verify the frequency accuracy of the analyzer over its entire operating
frequency range. A frequency counter is used to determine the analyzer's output frequency.
Analyzer warm-up time: 30 minutes
Specifications
Frequency Range
300 kHz to 3 GHz
b
30 kHz to 6 GHz
a. At 25 °C ±5 °C.
b. Only for analyzers with Option 006.
Required Equipment
Description
Frequency Counter
RF cable set, 50Ω type-N
2-way power splitter, 50Ω
Adapter, type-N (f) to BNC (m)
Procedure
1. Connect the equipment as shown in
2-4
a
Frequency Accuracy
±10 ppm
±10 ppm
HP/Agilent Part or Model Number
5350B/51B/52B
11851B
11667A Option 001
1250-0077
Figure
2-1.
Chapter 2

Advertisement

Table of Contents
loading

Table of Contents