Agilent Technologies 8753ES Option 011 Service Manual page 350

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Error Terms
Full Two-Port Error-Correction Procedure
Source Match (E
SF
Description Source match is a measure of test port connector match, as well as the
match between all components from the source to the test port. These are the forward and
reverse uncorrected source match terms of the driven port.
Significant System Components
• load calibration kit device
• open calibration kit device
• short calibration kit device
• bridge
• test port connectors
• bias tees
• step attenuator
• transfer switch
• test port cables
Affected Measurements Reflection and transmission measurements of highly
reflective devices are most affected by source match errors.
Figure 11-3
Typical E
11-10
and E
)
SR
/E
without and with Cables
SF
SR
Chapter 11