Inspect the Error Terms

Error terms are a measure of a "system": a network analyzer, calibration kit, and any
cables used. As required, refer to
• The specific measurement calibration procedure used to generate the error terms.
• The routines required to extract error terms from the instrument.
• Typical error term data.
Use
Table 9-1
to cross-reference error term data to system faults.
Table 9-1
Components Related to Specific Error Terms
Component
Calibration Kit
load
open/short
Test Set
connectors
bridge
bias tee
transfer switch
step attenuator
power splitter
Analyzer
sampler
A10 digital IF
test port connectors
External cables
Chapter 9
Chapter 11 , "Error Terms,"
Directivity
Source
Match
X
X
X
X
X
X
X
X
X
X
X
X
X
X
for the following:
Reflection
Isolation
Tracking
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Accessories Troubleshooting
Inspect the Error Terms
Load
Transmission
Match
Tracking
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X